PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Rev. 1-201
Model F53 Flow Monitor/Totalizer (universal-mount 1/2 DIN)
49
4.3 Configuring
CALCULATION
Measurement
☞
SET VARIABLE X
The analyzer can display a calculated measurement derived
from the measured flow rates or volumes of two different
sensors. First, assign which measured sensor signals are to
be variables X and Y. Then select the formula to apply to
them. To complete configuration, set the CALCULATION
measurement display format. Any analog output and/or
relay can represent the calculated measurement.
NOTE:
For calculated measurement, selected sensors must:
•
Be properly connected and operating.
•
Represent the same measurement (flow rate or
volume -- not one of each) and have the same
measurement units. However, the measuring
range for each sensor can be different.
1.
Press
MENU key
to display
&21),*85(
7(670$,17
(;,7
. Use
Ø
Ø
key
to select the “CONFIGURE” line.
2.
Press
ENTER key
to display
. Use
Ø
Ø
key
to select the “CALCULATION” line.
3.
Press
ENTER key
to display
.
4.
With the “SET VARIABLE X” line selected, press
ENTER key
to display a screen like
.
Use
Ø
Ø
and
×
×
keys
to view choices for VARIABLE X:
•
FLOW A
•
VOLUME A
•
FLOW B
•
VOLUME B
•
FLOW C
•
VOLUME C
•
FLOW D
•
VOLUME D
5.
With the desired choice displayed, press
ENTER key
to
enter this selection.