PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Model F53 Flow Monitor/Totalizer (universal-mount 1/2 DIN)
Rev. 1-201
60
4.6 Configuring Pulse
Outputs (A and B)
SET PARAMETER
(representation)
SET VOLUME
Increment
The analyzer provides two SPDT pulsed contact closure
outputs (A and B). Each pulsed output can be assigned to
represent a measured volume, and set to activate at a de-
sired volume increment for a preset time duration.
These
instructions configure Pulse A.
Configure Pulse B in the
same way using its respective menu screens.
1.
Press
MENU key
to display
&21),*85(
7(670$,17
(;,7
. Use
Ø
Ø
key
to select the “CONFIGURE” line.
2.
Press
ENTER key
to display
. Use
Ø
Ø
key
to select the “SET PULSE A” line.
3.
Press
ENTER key
to display
.
4.
With the “SET PARAMETER” line selected, press
ENTER key
to display a screen like
.
Use
Ø
Ø
and
×
×
keys
to view the choices (VOLUME A,
VOLUME B, VOLUME C or VOLUME D). With the de-
sired volume to represent the pulse output displayed,
press
ENTER key
to enter this selection.
Now set the volume increment at which you want the pulse
output signal to activate.
1.
After the
screen re-appears, use
Ø
Ø
key
to select the “SET VOLUME” line.