PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Rev. 1-201
Model F53 Flow Monitor/Totalizer (universal-mount 1/2 DIN)
55
☞
SET FILTER Time
Select SCALE 0 mA/
4 mA (low endpoint)
4.
Press
ENTER key
to display a screen like
. Use
arrow keys
to set the displayed
value at which 20 mA is desired, and press
ENTER key
to enter the value.
NOTE:
If the same values are set for 0/4 mA and 20 mA, the
output automatically goes to, and remains at, 20 mA.
A time constant (in seconds) can be set to filter or “smooth
out” the output signal. A minimum value of “0 seconds” has
no smoothing effect. A maximum value of “60 seconds” pro-
vides maximum smoothing. Deciding what output filter time
to use is a compromise. The higher the filter time, the longer
the output signal response time will be to a change in the
measured value.
1.
With the
screen displayed, use
Ø
Ø
key
to select the “SET FILTER” line.
2.
Press
ENTER key
to display a screen like
. Use
arrow keys
to adjust the dis-
played value to the desired filter time, and press
ENTER key
to enter the value.
Select each output to be 0-20 mA or 4-20 mA.
1.
With the
screen displayed, use
Ø
Ø
key
to select the “SCALE 0mA/4mA” line.
2.
Press
ENTER key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view both
choices (0 mA or 4 mA). With the desired choice dis-
played, press
ENTER key
to enter this selection.