PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Rev. 1-201
Model F53 Flow Monitor/Totalizer (universal-mount 1/2 DIN)
39
SET DECIMAL
(for displayed flow rate)
SET FILTER Time
WARNING:
CHANGING THE MULTIPLIER AUTOMATICALLY
REPLACES USER-ENTERED VALUES FOR THAT
SENSOR -- EXCEPT ITS CONFIGURATION AND
CALIBRATION DATA -- WITH FACTORY-DEFAULT
VALUES.
4.
With the desired choice displayed, press
ENTER key
to
enter this selection.
After choosing the multiplier, select the sensor flow rate to be
displayed as a whole number or decimal. The selected format
will also appear on all applicable edit/selection menu screens.
1.
With the
screen displayed, use
Ø
Ø
key
to select “SET DECIMAL” line.
2.
Press
ENTER key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view all choices
(XXXX, XXX.X, or XX.XX).
WARNING:
CHANGING THE DECIMAL POINT AUTOMAT-
ICALLY REPLACES USER-ENTERED VALUES FOR
THAT SENSOR -- EXCEPT ITS CONFIGURATION
AND CALIBRATION DATA -- WITH FACTORY-
DEFAULT VALUES.
3.
With the desired choice displayed, press
ENTER key
to
enter this selection.
A time constant (in seconds) can be set to filter or “smooth
out” the sensor signal. A minimum value of “0 seconds” has
no smoothing effect. A maximum value of “60 seconds” pro-
vides maximum smoothing. Deciding what sensor signal