40
pH Flo
w C
ell
pH Flow Cell
| VERITY
®
1810 Conductivity and pH Monitor
Setup
The pH Flow Cell requires some assembly.
There are four parts: the base, the O-ring, the pH sensor, and the securing nut.
When unpacking and storing the pH sensor, refer to the manufacturer’s instructions provided with the pH
sensor.
NOTE
The O-ring provided in the soaker bottle with the pH sensor is not required for
this setup and should remain in the soaker bottle cap for future storage of the pH
sensor.
After unpacking the items, do the following:
1. Place the O-ring on the black (thinner) end of the pH sensor.
2. Insert the black end of the pH sensor with O-ring into the base.
3. Pass the end of the pH sensor cable through the securing nut.
4. Hand-tighten the securing nut to the base.
NOTICE
Do not use tools.
Do not overtighten.
securing
nut
O-ring
(part number
5464549103)
base
Figure 17
pH Flow Cell (Exploded View)
pH
sensor
Figure 18
Soaker Bottle
O-ring