15
Technic
al Specific
ations
VERITY
®
1810 Conductivity and pH Monitor |
User’s Guide
VERITY® 1810 Conductivity and pH Monitor
Technical
Specification
Definition
Monitor Channels
Three: Conductivity, pH, and Temperature
pH Flow Cell
Accuracy
±0.1 units at 25°C
Maximum
Drift
0.1/10 hours at constant conditions
Part Number
11100001
Pressure
psi
bar
MPa
100
6.89
0.69
Readable
Range
0–14 pH
Specification
Range
2–12 pH
Wetted
Materials
Acrylic, PVDF, pH Glass, epoxy, BUNA-N
Volume
110 µL
Power
Requirements
USB powered instrument
Instrument Power Requirement: 5V DC (1W)
Recommended
Flow Rate Range
0.5–150 mL/min
Safety and
Compliance
The instrument has been certified to safety standards specified for Canada, Europe, and the
United States. Refer to the instrument rear panel label and the Declaration of Conformity
document for the current standards to which the instrument has been found compliant.
Software Control
PC control via USB and TRILUTION® LC Software
Weight
0.34 kg (0.75 lbs.)