41
pH Flo
w C
ell
VERITY
®
1810 Conductivity and pH Monitor |
User’s Guide
Plumbing Connections
Tubing and fittings for inlet and outlet tubing connections are ordered separately so as to be appropriate for
the application flow rate. The arrows on the base indicate the flow path.
Tubing and Fittings
Flow Rate (mL/min)
Part Number
Description
0.5–5
49951059
TUBING,PEEK .010IDx.063OD 5FT
5–20
49952059
TUBING,.020"X 1/16 PEEK 5 FT. LONG,ORA
20–50
49953059
TUBING,.030" X 5FT, PEEK
>50
49954059
TUBING,.040" X 5FT, PEEK
All Flow Rates
49041034
FERRULE, 1/16" ANTI-TWIST, PEEK (P-250)
All Flow Rates
490410255
NUT, 1/16" ANTI TWIST PEEK (P-255)
To make the plumbing connections:
1. On each end of the tubing, slide the
nut, then the steel compression
ring, and then the ferrule. The
flattened end of the ring should face
towards the nut with the narrow
end of the ferrule towards the ring.
2. Insert the nut at each end into its
port. Hold the tubing to the bottom
of the fitting while tightening the
nut to finger tight.
Figure 19
Plumbing Connections
The flattened end of the
ring should face towards
the nut
The narrow
end of the
ferrule is to
touch the ring
Figure 20
Plumbing Connections
inlet
outlet