18
Plumbing C
onnec
tions
Installation
| VERITY
®
1810 Conductivity and pH Monitor
Plumbing Connections
Inlet and outlet tubing are ordered separately so as to be appropriate for the application flow rate. Refer to
the
Replacement Parts and Accessories
appendix for part numbers.
Fittings are supplied.
To make the plumbing connections:
1. On each end of the tubing, slide the nut,
then the steel compression ring, and then
the ferrule. The flattened end of the ring
should face towards the nut with the narrow
end of the ferrule towards the ring.
2. Insert the nut at each end into its port. (Ports
are not labeled as the direction of the liquid
flow is not specific.) Hold the tubing to the
bottom of the fitting while tightening the
nut to fingertight.
NOTE
If the system includes a pH flow cell, one connection goes to the pH flow cell inlet. Refer to the
appendix for instructions and more information.
Figure 3
Plumbing Connections (Exploded View)
Figure 2
Plumbing Connections
The flattened end of
the ring should face
towards the nut
The narrow
end of the
ferrule is to
touch the ring