Conductivity Flow Cell
Calibration
It is recommended to calibrate the conductivity flow cell annually. Contact your local Gilson representative
for more information and assistance.
Replacement
Refer to the instructions in this section when replacing the
conductivity flow cell.
REMOVE INSTALLED FLOW CELL
1. Disconnect the USB cable from the back of the
instrument.
2. Disconnect the inlet and outlet tubing from the
flow cell.
3. Hold the installed flow cell securely with your hand,
and then lift the flow cell up and out. The flow cell is
secured with magnetic force.
INSTALL NEW FLOW CELL
1. Insert the new flow cell into its holder on top of the
monitor. It will only fit one way.
2. Connect the inlet and outlet tubing to the new
flow cell.
Figure 13
Conductivity Flow Cell Removed