Operation
Manual 20791, Rev. C, June 2001
2-19
[
ABS PWR PATHCAL]
Menu F1
: Toggles between absolute power measurements using Cal Factors, and relative measure-
ments that are being ratio’ed with a stored path cal.
When the Absolute Power measurement mode is active (ABSPWR shown on the menu in light
blue), the instrument applies a Cal Factor to each reading. All power readings are displayed in abso-
lute power units (dBm or Watts) unless the ratio of two sensors is being measured.
When Path Calibrations are used (PTH CAL shown on the menu in light blue), measurements are
ratio’ed to a stored path calibration. Cal Factors are not used.
Path calibrations can be used only when making swept measurements. If the entry channel is mea-
suring in the CW mode, the [
ABS PWR/PATH CAL
] softkey label will be in dim, dark blue and the
sensor path calibrations will have no effect on the CW measurements.
[
TRACE MEMORY]
Menu F1:
Accesses stored measurement traces. The entry channel current measurement trace can
be stored, the current measurement trace can be replaced with a stored trace, or the channel current
trace can be ratio’ed with a stored trace. This feature is available only when making swept measure-
ments.
When [
TRACE MEMORY
] is pressed, the softkey menu will change to menu F1-1.
[
MEAS-TRACE N]
Menu F1-1:
Allows you to specify any memorized trace that you wish to ratio with the current
trace.
The entry area of the display will present a prompt to enter the memory location number
(0 - 9) of the stored trace to be ratio’ed with the current measurement trace. A memorized trace
can only be entered through the keypad, not by the spin knob. When a UNITS key (MHz dBm
or GHz dB) is pressed, the measurement will ratio the current measurement trace with the
specified memorized trace.
The current measurement trace of any channel can be ratio’ed to only one memorized trace. If
the current measurement trace is being ratio’ed to one memorized trace and another memorized
trace is specified, then the current trace will be ratio’ed only to the last specified memorized
trace. Meas Trace x - Meas Trace y cannot be defined.
[
MEAS]
Menu F1-1:
Causes the entry channel to display its current measurement ratio’ed with or with-
out path calibration(s) as indicated, but not with any memorized traces.
[
TRACE
→
TRACE N]
Menu F1-1:
Memorizes the current measurement trace as Trace N. When this softkey is
pressed, a prompt will appear in the entry area of the display to remind the user to enter the
trace number under which the trace will be stored. When a UNITS key is pressed, the trace
will be stored in non-volatile memory. The entry area of the display will then indicate that the
trace was stored.
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...