Remote Operation
Manual 20791, Rev. C, June 2001
3-65
Modifier Description:
This modifier defines a SWEEP TIME for the start/stop
swept frequency mode of signal source. The query flag
(‘?’) affixed to the sweep time modifier queries the value
of the sweep time. If set, the next time the 8003 is talk-
addressed it will output the value of the sweep time in
ms. The resolution is 1 ms. The format of the returned
data is D, where D is two to five digits.
Modifier name:
SWPT, SWPT?, ST, ST?
Modifier Syntax:
PL[?]
[
ds
numerical value]
numerical value range:
Dependent upon the signal source if the source is
supported on the 8003’s private bus. If not supported
then the range is 99000 to 40 ms.
Modifier Description:
The update modifier decouples analyzer frequency
updates from sweeper frequency updates. This is most
commonly used when the analyzer and sweeper are set
to different frequencies, as in mixer or frequency
multiplier measurements. The analyzer normally uses
frequency information to change the operating frequen-
cies of the sweeper, annotate x-axis of the swept
display, apply correct Cal Factors to the reading, take
the correct portions of trace memories and path cal
memories, and fix cursors and markers. The update
modifier allows independent control of the sweeper
frequencies and all other frequency related analyzer
functions. If the modifier UPD ON is included once with
the SWPF, SWPS, or FIXED verbs, then all subsequent
uses of those verbs will update the sweeper frequency
along with the analyzer frequency. If the modifier UPD
OFF is included once with the above verbs, then all
subsequent uses of those verbs will change only
analyzer frequency parameters.
Modifier name:
UPD
Modifier syntax:
UPD
hs
state
State definition:
ON, OFF
Modifier Examples:
Send the following commands to modify both sweeper
and analyzer frequencies to 2000 to 4000 MHz.
SWPF;fa 2000;fb 10000;pl -10;st 500
SWPF;UPD ON
SWPF;START 2000;STOP 4000
Send the following commands to modify only the
analyzer frequency to 6000 to 8000 MHz
SWPF;UPD OFF
SWPF;START 6000;STOP 8000
Related Command:
SWP
Description:
This verb places the signal source mode to start/stop swept frequency. This is an instru-
ment wide command and, as such, has no channel or sensor specifiers. However there
are many parameters associated with the source mode. These are defined by the modifi-
ers which follow the verb.
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...