8003 Precision Scalar Analyzer
2-36
Manual 20791, Rev. C, June 2001
2.5.4
CAL Key
[CAL]
The FUNCTION
[CAL]
key is for path calibration, sensor calibration, and zeroing sensors and/or
bridges. When
[CAL]
is pressed, the softkey menu will be as shown in F4 in Figure 2-11.
[
SHORT/OPEN]
Menu F4:
Allows you to perform a short/open path calibration on a bridge, or to update the bridge
path calibration contained in memory.
When [
SHORT/OPEN
] is pressed, the softkey menu will change to allow the selection of input A,
B, or C for the short/open calibration.
See [
CLR PTH CAL
] in the following descriptions regarding clearing path calibrations.
The softkey display will change to Menu F4-1 to present the option to either store the short data or
return to Menu F4.
[
STORE SHORT (A, B, or C)]
Menu F4-1:
Initiates the collection of data which will be stored as the short path calibration
data. When the key associated with the sensor of interest is pressed, the softkey menu will go
blank. The active entry area of the display will contain the message, Storing Short. If the 8003
is controlling the source, it will temporarily select a source sweep time suitable for storing the
large number of data points needed (up to 4096). The short data will be stored in memory as
the short path calibration. Once this is done, the softkey menu will change to offer the option
of either going through the same procedure for the open condition and storing the open data or
returning to Menu F4.
Figure 2-11: Function Menus Accessed with the CAL Key
F4
F4-5
F4-4
F4-3
F4-2
F4-1
CAL
CAL
CAL
CAL
CAL
CAL
CAL
KEY
SHORT /
OPEN
THRU
CLEAR
PTH CAL
NF *
CAL
SENSOR
ZERO
SENSOR
ZERO ALL
SENSOR A
ZERO
SENSOR B
ZERO
SENSOR C
ZERO
RETURN
CAL ALL
SENSOR A
CAL
SENSOR B
CAL
SENSOR C
CAL
RETURN
CLR PTH
CAL A
CLR PTH
CAL B
CLR PTH
CAL C
RETURN
STORE
THRU A
STORE
THRU B
STORE
THRU C
A FULL
BAND Y / N
B FULL
BAND Y / N
C FULL
BAND Y / N
RETURN
STORE
SHORT A
STORE
SHORT B
STORE
SHORT C
A FULL
BAND Y / N
B FULL
BAND Y / N
C FULL
BAND Y / N
RETURN
NOTE:
The labels in the shaded boxes are for menu identification only
* Used only with Noise
Figure Testing.
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...