8003 Precision Scalar Analyzer
2-62
Manual 20791, Rev. C, June 2001
[
SET CAL S/N]
Menu SY1-14:
Defines the calibrator serial number. When this softkey is pressed, the pres-
ently defined calibrator serial number will flash and a prompt will be presented asking for a new
serial number to be entered as desired. Any number up to eight digits in length can then be
entered by using the 0 through 9 number keys
[
DEFINE PASSWORD]
Menu SY1-14:
Defines a password that users will have to enter to change the calibrator out-
put. When this softkey is pressed, a prompt is presented asking for a new password. Six number
keys can then be pressed. The keys being pressed will not be shown on the display. Once the six
number keys have been pressed, the display will again present a prompt asking that the pass-
word just defined be entered again to confirm the password. When the password is confirmed,
the defined password will be enabled.
[
CLEAR PASSWORD]
Menu SY1-14:
Clears the password. Once this softkey is pressed, the password will be cleared
and future users will be able to adjust the calibrator output without having to know a password
until one is again defined.
[
ABORT]
Menu SY1-14:
Abandons any changes made to the calibrator output adjustment. When this
softkey is pressed the routine will return to the SET CAL softkey of menu SY1-13. Any
changes that may have been made to the calibrator output and/or the calibrator serial number
and/or the password since entering the SET CAL menu will be reversed.
When the SET CAL softkey menu is displayed, it will only be possible to leave the menu and
return to regular data display by pressing the [
ABORT
] or [
DONE
] softkeys.
[
DONE]
Menu SY1-14:
Terminates the setting of the calibrator output. The calibrator output adjust-
ment and the calibrator serial number will be remembered as they appear at the time the soft-
key is pressed. The 8003 will also memorize the time when the DONE softkey is pressed as the
new calibration time. The instrument will return to the SET CAL softkey of menu SY1-13.
[
SET TIME]
Menu SY1-13:
The 8003 contains a real time clock with a battery backup so that it retains the
time and date at power off. The SET TIME softkey sets the clock time and date. When the
softkey is pressed, the regular data display will be replaced with a display of currently defined
time and date. Menu SY1-15 will also appear allowing you to set the hour, minute, second,
date, month and year. Setting of the hour, minute, and second of the clock is done using
24-hour time, 0:00:00 to 23:59:59. Years are defined by four digits.
NOTE:
Jumper W1 on the Calibrator PC Board must be moved to pins 2 and 3
from pins 1 and 2 to complete the enabling of password protection.
☛
☛
☛
☛
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...