Operation
Manual 20791, Rev. C, June 2001
2-91
2.11.4
Bandwidth Searches
The 8003 Analyzer can also find the bandwidth of frequency selective devices. Press
[CURSOR]
followed by the [
SEARCH
] softkey. Then press the [
BANDWIDTH
] softkey. The analyzer will
automatically find the maximum point in the pass band of the device, then find the -3 dB points on the
right and left of the maximum, and calculate the bandwidth.
Once the [
BANDWIDTH
] softkey is pressed, the search level can be changed by simply entering a new
value on the numeric entry keypad and pressing
[GHz/dB]
.
Figure 2-47: Automatic Bandwidth Movement (Very Narrow Bandpass Filter)
SEARCH
RIGHT
BAND-
WIDTH
SEARCH
LEFT
SEARCH
VALUE
CURSOR
RETURN
STRT
1.711 GHz
STOP
SRCH
-3.00
dB
7.193 GHz
CH1: SW A -PC
CSR
-0.00 dB
5.0
dB/ H
REF -0.00 dB
∆
20.000 GHz
MAX
1
∆
CSR
7.193 GHz
MIN
VALUE FOUND: TRACE HOLD
+
∆
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...