8003 Precision Scalar Analyzer
2-22
Manual 20791, Rev. C, June 2001
2.5.2
SCALE Key
[
SCALE]
The FUNCTION
[
SCALE]
key alters the parameters related to the graphical presentation of the
measured data. These parameters include display scaling, display reference level, and display position
Scaling functions are available only for swept measurements, and will be shown as unavailable (in dim,
dark blue on the softkey menu display) if the active channel is in the CW mode.
[
SCALE FACTOR]
Menu F2:
Allows the setting of the scale factor for the active entry trace. It causes the entry area of
the display to indicate the current scaling factor for the entry channel, and also presents a prompt
to enter a new scaling factor, if required. If the entry channel is continuously autoscaling when this
softkey is pressed, the channel will cease autoscaling and the last scale factor determined by the
autoscaling will be the scale factor shown in the entry area of the display.
Scale factors can be either selected by the spin knob or entered through the numeric keypad. If
numerically entered, the scale factor will always be rounded to the nearest allowable value. The
allowable scale factors are: 0.1, 0.2, 0.5, 1, 2, 5, 10, and 20 dB/DIV.
[
AUTOSCALE]
Menu F2:
Causes the 8003 to autoscale the entry channel. Autoscale involves setting the scale fac-
tor, the reference level, and the reference position to display the entry channel main trace with the
highest possible resolution while still fitting the entire trace on the display. When [
AUTOSCALE
]
is pressed the 8003 autoscales the entry channel, the parameters are changed, the display adjusts,
and the entry area displays the main trace scaling factor so that it can be changed, if desired.
[
CONT AUTOSCL]
Menu F2:
Causes the 8003 to begin continuously autoscaling the entry channel. Continuous
autoscaling means that the routine will try to maintain the greatest resolution possible while still
fitting the entire trace in the display. During every update of the channel’s trace, the 8003 will
internally autoscale the channel. The scaling that results from this autoscale will be compared with
both the channel current scaling and the results from recent autoscales. Using hysteresis, the chan-
nel will gradually change the scaling on the channel as required.
[
REF POS]
Menu F2:
Sets the position of the reference line display graticule for the entry channel main trace.
The current reference position is indicated by a small number, corresponding to the channel num-
ber, to the left of the grid. The reference position can only be changed by using the spin knob.
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...