2.9.2
Two-Sensor Insertion Loss Measurements
The basic two-sensor insertion loss measurement consists of monitoring the power incident upon the
DUT, and comparing it to the power exiting the DUT. Incident power can be monitored in a variety of
ways, but the two most common are to use a Power Splitter or Directional Coupler
,
shown in
Power splitters have the advantage of wide frequency range, typically from dc to tens of Gigahertz.
However, power splitters typically have an insertion loss of at least 6 dB. Directional couplers with an
insertion loss typically less than 1 dB are used when insertion loss is a concern. However, wide frequency
coverage is often more difficult and expensive than with power splitters.
Figure 2-31: Two Sensor Insertion Loss Setup w/ a Power Splitter
Figure 2-32: Two Sensor Insertion Loss using a Directional Coupler
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...