Operation
Manual 20791, Rev. C, June 2001
2-57
2.7
SYSTEM Controls
The system controls are keys that affect the instrument as a whole. This includes the keys used to
configure the instrument and to preset the instrument. The SYSTEM section of the 8003 consists of the
[CONFIG]
and the
[PRESET]
keys.
2.7.1
PRESET Key
[PRESET]
The PRESET function key changes parameters that have been set for currently displayed back to the
factory-set defaults. See Appendix B for a listing of the 8003 factory-set default values.
2.7.2
CONFIG Key
[CONFIG]
The sub-menus accessed through the
[CONFIG]
key contain a softkey label called either [
RETURN
]
or [
DONE
] located at the bottom of the menus. Pressing this softkey will always return the routine to
the menu immediately preceding the menu currently being displayed. Pressing the softkey more than
once allows you to step backward through the menu structure to a preceding menu.
The CONFIG function key configures the scalar analyzer and any other devices that might be on its
private GPIB bus. When
[CONFIG]
is pressed, menu SY1 shown in Figures 2-16 and 2-17 (pages 2-58
and 2-59) will appear.
[
SENSORS AC/DC]
Menu SY1:
Changes the detection mode of the sensors. Pressing this softkey will bring up menu
SY1-1 to select either all or individual sensors as AC or DC. The state (AC or DC) that the setup
sensor is currently in will be displayed in light blue. The other state that the sensor would enter if
the softkey were pressed is displayed in standard menu green.
[
SENSORS CW FREQ]
Menu SY1:
Provides access to a series of sub-menus that allow you to specify which sensors will
take measurements in the CW mode, and what frequency correction factors will be used with the
sensors.
Whenever a channel is taking measurements in the swept mode, the 8003 determines the frequency
of the signal from the sweep input. The instrument will expect the sweep input to be sweeping from
0 V to 10 V, with 0 V representing the start frequency and 10 V representing the stop frequency.
The start and stop frequencies are assumed to be those that were specified by the use of the source
control keys. The 8003 will automatically determine the frequency of the signal from the sweep
input and correct the power readings of the sensor according to the frequency correction table
stored in the sensor’s PROM. See also the START key on page 2-52 and the STOP key on page
2-53 for the setting of the start and stop frequencies.
When a sensor is measuring in the CW mode it is possible to identify the frequency directly, specify
a cal factor, or command the 8003 to automatically determine the frequency of the signal from its
control of the source. If the 8003 is commanded to automatically determine the frequency of the
signal, it will use the source’s CW output frequency that was set by the 8003. If the 8003 is not con-
trolling the source, the 8003 will still use the frequency set in the source control section. The speci-
fication of the CW mode of frequency correction is accessed through the [
SENSOR CW FREQ
]
softkey.
The softkey label that represents the currently defined mode of CW frequency correction for the
setup sensor will be light blue. All other softkey labels will be in standard menu green.
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...