8003 Precision Scalar Analyzer
3-44
Manual 20791, Rev. C, June 2001
650
ENTER 704;Reading! In free running mode, analyzer gives last readin
660
! any time that it is talk-addressed.
670
Counter=1
680
END WHILE
690
Tottime=TIMEDATE-Starttime
700
OUTPUT 704;PWR 1;T0! Good housekeeping returns to display update mode
710
WAIT 1! Always wait after mode change
720
!
730
PRINT Free running mode msmts/second = ;Maxcount/Tottime,Power = ;Rea
740
!
750
!.............................................................
760
! MEASUREMENT OF TRIGGERED FAST CW MODE MEASUREMENT SPEED
770
!
780
OUTPUT 704;PWR 1;T2! First set triggered fast cw mode = T2
790
WAIT 1! IMPORTANT** wait for analyzer to change modes and update
800
! temperature correction
810
!
820
! Initialize loop variables
830
Maxcount=1000! Take 1000 readings
840
Counter=0 ! Initialize counter
850
!
860
! MAIN LOOP
870
Starttime=TIMEDATE
880
WHILE Counter<Maxcount
890
TRIGGER 704! Triggered mode requires Group Execute Trigger
900
ENTER 704;Reading
910
Counter=1
920
END WHILE
930
Tottime=TIMEDATE-Starttime
940
!
950
PRINT Triggered mode msmts/second = ;Maxcount/Tottime,Power = ;Readin
960
!
970
OUTPUT 704;PWR 1;T0! Good housekeeping to return analyzer to
980
! standard cw mode to update display
990
WAIT 1
1000 !
1010 !.............................................................
1020 ! EXAMPLE OF 2 CHANNEL MEASUREMENT IN FAST-TRIGGERED MODE
1030 !
1040 OUTPUT 704;POWER 1,A;ON;POWER 2,A;ON;CHAN 3;OFF;CHAN 4;OFF
1050 ! Note that both channels are reading the same sensor A. This is just
1060 ! for convenience in running a demo. It is just as easy to define
1070 ! channel 2 to read sensor B. With 2 different sensors, the msmt speed
1080 ! is almost identical to the code running here.
1090 !
1100 OUTPUT 704;PWR 1;T2! First set triggered fast cw mode = T2
1110 WAIT 1! IMPORTANT** wait for analyzer to change modes and update
1120 ! temperature correction
1130 !
1140 ! Initialize loop variables
1150 Maxcount=1000! Take 1000 readings
1160 Counter=0 ! Initialize counter
1170 !
1180 ! MAIN LOOP
1190 Starttime=TIMEDATE
1200 WHILE Counter<Maxcount
1210
TRIGGER 704! Triggered mode requires Group Execute Trigger
1220
ENTER 704;Reading a,Reading b
1230 ! Now, analyzer returns 2 readings for every trigger
1240
Counter=1
1250 END WHILE
1260 Tottime=TIMEDATE-Starttime
1270 !
1280 PRINT Sensor A = ;Reading a;, Sensor B = ;Reading b
1290 PRINT 2-channel msmt speed = ;Maxcount/Tottime; rdgs/second
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...