8003 Precision Scalar Analyzer
2-14
Manual 20791, Rev. C, June 2001
When path calibration is used, the 8003 does not apply Cal Factors to the readings. Cal Factors are only
useful for absolute power measurements. Any Cal Factor correction would be canceled when making
relative measurements.
To make the path calibration over wide frequency ranges useful, the 8003 memorizes a very large
number of data points (up to 4096) during a path calibration. With such a high resolution of data
points, the 8003 is able to maintain a very high level of accuracy within any sub-range of the frequency
range of the path calibration.
Any path calibration can be cleared at any time during instrument operation. When a path calibration
is cleared, it is initialized to 0 dBm so that it will have no effect if it were ratio’ed with any
measurements.
2.3.7
Trace Memory
The 8003 can store measurement data of up to ten traces in non-volatile memory. The traces are stored
and used by number. The traces can be stored by any channel and they, in turn, can be used by any
channel. The measurement can be specified on any desired channel to be the ratio of the current
measurement data and the trace n measurement data, where n is between 0 and 9 inclusive. Traces are
memorized as 512 points along the frequency axis.
A title can be specified for each stored trace. This allows you to specify in your own words what the
memorized trace data represents. When a measurement is being ratio’ed with a memorized trace, the
title of the trace memory replaces the channel title. Memorized traces can be displayed as a channel for
comparison to a current measurement.
A memorized trace can be deleted by clearing all of the memory of the 8003, by clearing specified
memorized traces, or by clearing all memorized traces. When a memorized trace is cleared, it will be
replaced with a 0 dBm trace, so that it would have no effect if it were inadvertently ratio’ed with a
measurement trace.
2.3.8
Power On
When the 8003 is powered on, the instrument will restore itself to the same configuration it was in at
power off. This will include restoring the menu which was being displayed at power off and the path
calibrations at the time power was turned off. The only aspects of the instrument operation which will
not be restored will be the measurement data at the time power is turned off.
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...