8003 Precision Scalar Analyzer
3-24
Manual 20791, Rev. C, June 2001
3.3.16
Input Command Structure
INPUT
Description:
This verb inputs frequency, trace, pathcal, and cal factor information to the 8003 for use in
specific test routines. When information is downloaded, the 8003 needs to know the span
of the frequency range that the data represents. The first two values expected by the
analyzer, therefore, will be the Start and Stop frequencies. The ensuing data will then
represent values of power evenly spaced over the specified range. The first data item is a
power reading at the start frequency, and the last data item is a power reading at the stop
frequency. 512 power values are expected for trace memory, while 4096 values are
required for sensor Pathcal and Cal Factor data. The units for frequency are MHz, and the
units for power are dB. All data must be separated by a comma and end with a line feed.
Syntax:
INPUT
us
memory modifier
hs
sensor/trace specifier
Modifiers:
Trace specifiers:
0 through 9
Sensor specifiers:
A, B, C
Pathcal specifiers:
A, B, C
Calfactor specifiers:
A, B, C
Trace:
This refers to the 10 trace memories that can be used
for ratio measurements with any channel (see Section
2.3.7 on page 2-14 in the OPERATION Chapter of this
manual for a complete description).
Pathcal:
This refers to the sensor’s path calibration memory (see
Section 2.8.2 under
Frequency Response Correction
for
a complete description).
Calfactor:
This refers to the sensor’s calibration factor data (see
the paragraph starting at the bottom of Section 2.12 on
page 2-95 in the OPERATION chapter of this manual for
a complete description).
Data Input Format:
START FREQ,STOP FREQ
(MHz)
D.ddd
then 512 data points for
TRACE
(dB)
±
ddd.dd
then 4096 data points for
PATHCAL
±
ddd.dd
then 4096 data points for
CALFACTOR
±
ddd.dd
D = 0 to 6 digits
Example:
start frequency,stop frequency,data,(start freq),...,data(stop freq)
Immediately following the
INPUT
command string, the analyzer expects the data with the
above format.
Syntax Examples:
INPUT,TRACE 4
INPUT;PATHCAL B
INPUT;CALFACTOR C
Programming Examples:for HP 200 or 300 series controllers:
OUTPUT 704;INPUT;TRACE 4,Start_freq,stop freq,Data_array(*)
Related Command:
OUTPUT
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...