Operation
Manual 20791, Rev. C, June 2001
2-53
the softkey (START, CENTER or CW) will be displayed in the entry area of the display. For exam-
ple, if the source were in the Start/Stop mode, the START label would be in light blue. If the
[
CENTER
] softkey were pressed, then the CENTER label would change to light blue, the START
label would change to standard menu green, and the source would begin measuring in the Center/
∆
F mode using the frequencies that were last entered for the Center and
∆
F frequencies.
The frequency ranges that have been defined for the start/stop frequencies and the center/
∆
F fre-
quencies are completely independent. Whenever a change is made from one frequency mode (start/
stop, or center/
∆
F) to the other, the instrument will always revert to the frequency range that was
last defined for the mode to which the instrument is switched. Similarly the frequency of the source
CW output is also independent of the start, stop, center, and
∆
F frequencies. New frequencies are
entered using the numeric keypad and the appropriate terminator key.
2.6.4.3
STOP Key
[STOP]
The SOURCE
[STOP]
key initiates the source frequency changes. It allows the stop,
∆
F, or CW
frequencies to be specified depending upon the mode of operation. See the description of the UPD SWP
ON/OFF softkey section .
When
[STOP]
is pressed, the entry area of the display will show the current stop,
∆
F, or CW
frequency, depending upon whether the source is in the start/stop, center/
∆
F, or CW mode, respectively.
A prompt will be presented to change the frequency if desired, and the softkey menu will change to the
menu in Figure 2-14.
[
STOP], [DF] and [CW]
These softkeys change the mode and frequency of the source. The softkey label that corresponds to
the mode currently being used (STOP for Start/Stop,
∆
F for Center/
∆
F, and CW) will be in light
blue. The other two options will be in standard menu green.
Pressing one of the softkeys with a standard menu green label will cause the source to immediately
switch to measuring in the mode indicated by the softkey pressed. The current value for the fre-
quency associated with the softkey pressed (STOP,
∆
F, or CW) will be displayed in the entry area of
the display. For example, if the source were in the Start/Stop mode, then the Stop label would be in
light blue. If the
∆
F softkey were pressed, then the
∆
F label would change to light blue, the Stop
label would change to standard menu green, and the source would begin measuring in the Center/
∆
F mode using the frequencies last entered for the Center and
∆
F frequencies.
Figure 2-14: Stop Key Softkey Menu Prompt
STOP
STOP
F
CW
UPD SWP
ON/OFF
∆
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...