Operation
Manual 20791, Rev. C, June 2001
2-31
[
LIMIT LINES]
Menu F3-9:
Allows the limit lines for the entry channel to be edited or turned on and off. If the
channel is within the limits that have been set for it, the channel number will be displayed with the
word
PASS
. If the channel parameters are not within the set limits, the word
FAIL
will be dis-
played.
[
LMT LNS ON/OFF]
Menu F3-9:
Toggles the limit lines ON or OFF.
[
DEFINE SEGMENT]
Menu F3-9:
Generates a table on the screen for the entry of limit line parameters, and will also
cause Menu F3-10 to be displayed. First, it is necessary to select the type of limit line from
Menu F3-10. The choices are a POINT LIMIT, a FLAT LIMIT, and a SLOPE LIMIT.
When the softkey for the type of limit line is pressed, the 8003 will automatically assign a num-
ber between 1 and 12 to the segment to be defined and position a box to the START-FREQ
column in the Limit Lines Table. The frequency is entered using the numeric keypad, and the
digits will be shown in the data entry section of the display.
Appropriate values should also be entered for the remaining columns in the table depending on
the type of Limit Lines selected.
When all of the values for a given segment have been entered, the ABORT choice on Menu
F3-10 will change to DONE. Additional segments can be defined by assigning a new segment
number and pressing the softkey for the type of Limit Line desired.
When [
DONE
] on Menu F3-10 is pressed, you will be returned to Menu F3-9 at which stage
[
LMT LNS ON/OFF
] should be pressed to turn the Limit Lines ON. This will also activate the
PASS/FAIL flag on the display.
[
SEGMENT N CLEAR]
Menu F3-9:
Clears all data for the segment selected. Clearing a segment does not change the
value or position of any other data in the table.
[
CLEAR ALL SEGMENTS]
Menu F3-9:
Clears all of the segments for the active channel. It does not clear segments on
non-active channels.
[
PLOT ENTRIES]
Menu F3-9:
Makes a hardcopy of the user-entered segment data that defined the limit lines.
[
ABORT PLOT]
Menu F3-9:
Stops the printing process at any time as desired.
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...