Genevac HT-8 & HT-12 Series II Evaporating Systems
Page 60 of
64
04-4541
Issue 1-9 – June 2008
Appendix A
Programming Guide for Drying HPLC Fractions
Drying HPLC Fractions
Run number
1 2 3
Run Data Locked
N N N
Next Run Number
2 3 0
Run Name
Fractions
Fractions part 2
Fractions part 3
Sample Holder Type
Your Holder
Your Holder
Your Holder
Sample Volume ml
Your Volume
Your Volume
Your Volume
Solvent
MeCN or MeOH & H
2
0 MeCN or MeOH & H
2
0 MeCN or MeOH & H
2
0
SampleGuard Control Temp C
40 40 40
SampleGuard Control Channel
1 1 1
CoolHeat Enable Pressure
50 50 50
Chamber Temperature control
preheat to 25
preheat to 30
preheat to 30
Rotor Speed
High High Low
Heat-Off Option
Elapsed Time
Heat Flow
Elapsed Time
Heat-Off Elapsed Time
ask for time
-
01:00 - 03:00
End-of-Run Option
Elapsed Time
-
Elapsed Time
End-of-Run Elapsed Time
ask for time
-
01:00 - 03:00
Auto Defrost & Drain
Y N Y
Pressure Control Regime
Variable Dri-Pure
Controlled Pressure
Full Vacuum
Dri-Pure Start Pressure (mbar)
175 -
-
Dri-Pure End Pressure (mbar)
40 for MeCN
45 for MeOH
-
-
Dri-Pure Ramp Time
20 - -
Final Control Pressure
40 for MeCN
45 for MeOH
8 -
Notes
Remove organic phase Remove most of the
aqueous phase
Remove final
“stubborn” solvents
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com