Genevac HT-8 & HT-12 Series II Evaporating Systems
04-4541
Issue 1-9 – June 2008
Page 23 of
64
Rotor Speed
There are now 3 options
available:
High
Rotor speed applies
nominally 500 G.
Low
Rotor speed applies
nominally 300 G.
Very low
Rotor speed = 500 rpm,
applies nominally 50 G.
SampleGuard and
Coolheat are not
available.
Heat-Off Option
There are now 2 new
options available:
Heat Flow
Automatically detects
dryness by monitoring
the rate at which the
solvent is evaporating.
Low solvent loading or if
the lamps are turned off
may cause Heat Flow to
fail.
If failure occurs, select
one of the other two
options.
Sample Detection
Monitors the
temperature difference
between the 2
SampleGuard channels.
Notes on the use of optional end of run Automatic Defrost and Drain can be found in item 13, Annex A; Getting
the best from Auto Defrost Drain.