Genevac HT-8 & HT-12 Series II Evaporating Systems
04-4541
Issue 1-9 – June 2008
Page 37 of
64
Dispense the DMSO into the vials and load the
sample holders into the sample swings.
Always use the upper level when loading partial rotor
loads.
Always ensure that the vials, sample holders and
sample swings are correctly and securely located.
Position the SampleGuard probe in its’ location hole
in the sample holder or swing and connect to
channel 1 on the SampleGuard.
Position the sample SampleGuard probe in an outer
tube or well and connect it to channel 2 on the
SampleGuard (refer also to additional notes on page
26).
Always ensure that the tip of the sample probe is
located at the bottom of the tube or well.
Close and hold the door firmly against the door seal
and press and hold the
Door Close
switch until the
system beeps.
Press the
START
key.
The system will indicate that the chamber is warming
up.
This will take approximately twenty minutes.
When the chamber is up to temperature, the system
will start automatically.
With aluminium sample holders of this thermal
capacity, some time can be gained by preheating
them in an oven to 40° C.