Genevac HT-8 & HT-12 Series II Evaporating Systems
04-4541
Issue 1-9 – June 2008
Page 47 of
64
7.5
Moving your system
These notes are included to assist you in moving
your system to another area.
Before moving your system, the following key points
should be addressed.
•
Ensure that there is sufficient bench or fume
cupboard space to position your system.
•
Ensure that there is an adequate power supply.
•
Ensure that there is adequate ventilation.
•
Ensure that there is sufficient clearance for
adequate ventilation of the pump and condenser.
•
Ensure that provision has been made for the
drainage of condenser solvents and the pump
exhaust.
•
Ensure that the pump fluid is drained if the
system is to be transported any appreciable
distance (CVP only).
•
Ensure that the condenser pots are drained and
flushed.
•
Ensure that the sample swings are removed
from the rotor.
•
Ensure that all the clamps, seals, tubes, cables
and these instructions
are retained together.
Having moved your system, the following notes will
aid you to reconnect the system.
Connect the control lead from the pump to the
chamber.
There are two location blades that need to be
aligned with the socket.
Ensure that the orientation of the plug to the pump is
correct.