Genevac HT-8 & HT-12 Series II Evaporating Systems
Page 24 of
64
04-4541
Issue 1-9 – June 2008
During a run
The main section of the
RUN LOG
screen, gives a
graphical display of the real time changes in
chamber pressure, spin rate, sample and sample
holder temperatures.
The top left hand section, displays the current
temperatures of channels 1 and 2, the chamber
temperature, the condenser temperature, the
chamber pressure, the total time elapsed and the
stage time elapsed if linked runs are being
undertaken.
NOTE THE NEW GRAPHICAL
DISPLAY.
•
The RUN LOG screen
displays the real time
events.
•
The top left block gives
current values.
•
The top right block displays
any alarms.
•
The block at the foot of the
screen displays the current
status of the system and
Instructions.
•
Pressure is plotted on log
scale.
•
Press STOP to terminate a
run at any time.
Run finished press < > to view data.
any other key to end.
When run enters final drying
period the display changes to
indicate the time left to
End of
Run
or if linked runs
End of
Stage
.
Press the
DISPLAY
key during a
run to return to the
RUN LOG
screen.
Operating the
STOP
key at any
time during a run will terminate
the run.
The chamber will vent and the
rotor will start to spin down.
You will not be able to open the
chamber until the chamber has
reached atmospheric pressure
and the rotor has stopped
spinning.
When the chamber has vented
and the rotor stopped, you will be
given this
STATUS
prompt.