Genevac HT-8 & HT-12 Series II Evaporating Systems
Page 16 of
64
04-4541
Issue 1-9 – June 2008
Scroll
CVP
5.1 How to switch on the pump
Connect the pump to the mains and switch on the
mains.
Switch the pump mains switch on.
The pump will not be ready for use until the green
ready light on the front panel is illuminated (Approx 7
minutes) – but power is available to the evaporator.
The system will not be available for use until the
vacuum pump has reached the correct operating
temperature (about 25 – 30 minutes for the Scroll
pump. This only occurs when the system is first
switched on or when the Auto Stop function has
been used), (5 – 10 minutes for the CVP).
5.2 How to switch on the evaporator
Switch the mains switch on.
The screen will display the
Software Control
screen.
Press the
START
key on the keyboard to access the
Select Run
screen.
The
Select Run
screen will load up at the last run
that was used.
A series of runs have been preloaded from RUN
51 to RUN 100 inclusive.
Runs 51 to 98 provide a range of Standard
Configurations.
Runs 99 and 100 are only for use by engineers for
Maintenance/Diagnostics.
The condenser will be powered up when the
chamber is switched on.
There will be a slight delay until the condenser
reaches the correct temperature.