Genevac HT-8 & HT-12 Series II Evaporating Systems
04-4541
Issue 1-9 – June 2008
Page 9 of
64
IF THIS POSITIONING REQUIREMENT IS
IMPRACTICAL THEN THE EVAPORATOR
SHOULD BE BOLTED TO THE BENCH OR
TROLLEY, BY ITS FIVE MOUNTING FEET, USING
M10 H.T. BOLTS ZINC.
GENEVAC SHOULD BE CONSULTED FOR
ADVISE ON ANY OTHER POSITIONAL
REQUIREMENTS.
3.5 Fitting
the
vacuum
pump
One of the following pumps will
have been supplied with your
system:
Scroll:
The Scroll pump supplied by
Genevac is only available as a
Non Recirculating model.
CVP:
The CVP 100 pump as
previously supplied by Genevac
was available in two formats:
the standard (Non
Recirculating) model or the
optional Recirculating model.
General
The Genevac pump supplied
with your evaporator will have
the connecting pipe/pipes,
fittings and power control lead
included with it.
Connect your vacuum pump to
the evaporator using the
connecting pipe and flange
fittings.
Connect the power and control
lead between the pump and the
socket on the rear of the
chamber.
If your system has been supplied with a Scroll pump then it must not be moved
between systems without first consulting Genevac Service for advise.
The Scroll pump will only function correctly on a Genevac Evaporator that has
been upgraded to work with the Scroll pump.
If your system has been supplied with a CVP pump then it must have an adequate supply of cooling
air and the hot outlet must have at least 300 mm of space beyond the hot air grille in the pump base.
It should not be placed in a cupboard without special precautions to ensure adequate cooling.
Consult Genevac Service for advice in such cases.