SampleGuard Control Temperature
NOT AVAILABLE FOR LYOPHILISATION
(Very Low rotor speed)
Sample Guard Control Channel
NOT AVAILABLE FOR LYOPHILISATION
(Very Low rotor speed)
Coolheat Enable Pressure
NOT AVAILABLE FOR LYOPHILISATION
(Very Low rotor speed)
Chamber Temperature Control
Select
Done
and press
ENTER
.
Press
ENTER
and input the temperature you wish to
limit the sample to during evaporation.
This can be in the range of -20 to +70° C.
Note that if not specified, the default setting is 30° C.
Note this is not available for
Lyophilisation
.
Press
ENTER
and input the channel on which you
want to control temperature.
Channel 1 will normally be assigned to the probe in
the sample holder and channel 2 to the probe in the
sample.
Note that when powered down, this setting defaults
to channel 1 as the control channel.
Note this is not available for
Lyophilisation
.
Press
ENTER
and input the pressure below which
you want the lamps to be turned on.
This can be in the range of 5-400 mbar.
If nothing is selected, the setting will default to 400
mbar.
Note this is not available for
Lyophilisation
.
Press
ENTER
and Select the method from the 3
options displayed:
Wait for Chamber to Heat
Range from 0°C to 45°C.
Wait for Chamber to Cool
Range from 0°C to 45°C.
Automatic Control
Temperature is automatically set to an optimised
figure for the recognised solvent / mixture selected.
Input the minimum temperature you want the
chamber to reach before spin-up will commence.