Mechanical Separation /
GEA Westfalia Separator
4.29 Alarm #42 - SMS sensor fault (option)
SMS – SENSOR FAULT
During SMS CHECK
During TIMER 44
Alarm#42 ENTER
Text display
Possible causes
Action
PT2 no pressure signal.
•
Check setpoint PT2.
Filling water valve V2 no function.
•
Check function V2.
T44 too short.
•
Check T44.
Solids space blocked
Pressure sensor line to PT2.
•
Clean bowl, check bore in centri-
petal pump chamber cover.
Bowl not closed.
•
Check bowl hydraulics.
Filling water capacity during T44 is too
low.
•
Check water capacity (0.1/s)
Pressure switch PT2 defective.
•
Check pressure switch PT2.
Incorrect switching point at PT2.
•
Check switching point PT2.
•
Deselect function if required.
Help text – information on the control system
E N T E R
Step 1
Acknowledge fault
H E L P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
SMS SENSOR FAULT
Please check:
Setpoint PT2
Function V2
->
Page 2
Timer 44
Sludge space
Bore in centripetal pump
chamber cover
->
Page 3
Bowl hydraulics
Function PT2
back with ENTER
F 1