Mechanical Separation /
GEA Westfalia Separator
4.9
Alarm #11 - Vibration stage 1 (option)
VIBRATION STAGE 1:
Motor off after T25
Alarm#11 ENTER
Text display
Possible causes
Action
Limit value 1 longer than T25.
•
Check T25.
Uneven solid desposits in the bowl.
•
Check solids space and polyamide
gasket of the bowl top.
Vibo sensor has tripped.
•
Check sensor fastening.
Limit value 1 set too low.
•
Check setting.
Bowl not correctly mounted or wrong
parts from other bowl.
•
Open and clean the bowl and mount
in accordance with the manual.
CAUTION!
•
Use only parts with identical
serial number.
External vibrations
•
Check foundation fastening ar-
rangement.
Help text – information on the control system
E N T E R
Step 1
Acknowledge fault
H E L P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
VIBRATION STAGE 1:
Please check:
Timer 25
Solids space
->
Page 2
Polyamide gasket
Sensor fastening
Limit value 1
Correct bowl parts
->
Page 3
Foundation fastening
Back with ENTER
F 1