Mechanical Separation /
GEA Westfalia Separator
4.21 Alarm #34 - Bowl monitoring (option)
BOWL MONITORING
Speed too high
during TIMER 06
Alarm#34 ENTER
Text display
Possible causes
Action
Closing water pulse T06 too long,
bowl has opened.
•
Check T06.
Water pressure for hydraulics during
T06 too high, bowl has opened.
•
Check water pressure with open
valve V3.
Valve V3 is soiled, catches, remains
open too long.
•
Check valve V3.
New clutch shoes fitted.
•
Check speed offset.
New drive belt fitted.
•
Check drive belt.
•
Deselect function if required.
Note:
Alarm #34 only active when speed pickup is installed!