8555-9201-000
/
0109
49
Mechanical Separation /
GEA Westfalia Separator
4.4
Alarm #06 - TAH product feed
TAH PRODUCT FEED
Temp. >max
Separation - Stop
Alarm#06 ENTER
Text display
Possible causes
Action
Product too hot.
•
Check control valve.
Feed capacity too low.
•
Check feed capacity.
Setpoint TAH set too low.
•
Check setpoint TAH.
Controller setting or function not o.k.
•
Check controller setting and func-
tion.
Line compensation at controller not
correctly adjusted.
•
Check line compensation at control-
ler.
Temperature controller defective.
•
Check temperature controller.
PT 100 sensor defective.
•
Check PT 100 sensor.
Help text – Information on the control system
E N T E R
Step 1
Acknowledge fault
H E L P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
PRODUCT TEMP. MAX
Please check:
Control valve
Feed capacity
->
Page 2
Setpoint TAH
Switching stages electr.
Temperature controller
PT100 sensor
->
Seite 3
Back with ENTER
F 1