Mechanical Separation /
GEA Westfalia Separator
4.5
Alarm #07 - Ejection monitoring
EJECTION MONITORING
Eject. curr.< offset
Separation “Stop”
Alarm#07 ENTER
Textdisplay
Possible causes
Action
Operating water valve V3 is defective.
•
Check bowl hydraulic function with
operating water valve V3.
Connector on operating water valve V3
is loose.
•
Check connector on operating
water valve V3.
Offset setting for ejection current ad-
justed too high.
•
Check offset setting.
Water volume at V3 too low.
•
Check water volume.
Monitoring time T28 too short.
•
Check T28.
Current evaluation defective.
•
Check current transformer electri-
callly.
Filter insert in strainer at solenoid block
soiled.
•
Check filter insert.
Clutch shoes worn or oily.
•
Check clutch shoes.
Help text – Information on the control system
E N T E R
Step 1
Acknowledge fault.
H E L P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
EJECTION MONITORING
Please check:
Bowl function
Hydraulic with V3
->
Seite 2
Plug-in valvel V3
Current offset
Setting
Water volume
->
Page 3
T28
Current transformer electr.
Prestrainer
Back with ENTER
F 1