Chapter 10 - Care & Maintenance
10-21
10-7-7
Isolated Patient Lead (Source) Leakage–Lead to Lead
When using the Dale 600, switch the meter's function selector to the LEAD-LEAD position.
Select and test each of the ECG lead positions (except ALL) on the LEAD selector, testing each to the
power and ground condition combinations found in:
•
•
, as applicable.
Record the highest leakage current measured.
10-7-7-1
Lead to Lead Leakage Test Record
shows a typical format for recording the patient lead to lead leakage current.
Measurements should be recorded from each lead combination under each set of test conditions
specified in:
•
•
, as applicable.
Record all data on the EQC inspection certificate.
Table 10-18 Typical Data Format for Recording Patient Lead to Lead Leakage
System Under Test:
________________________
Date of Test:
_____________________
NOTE:
Values in italics font are given as examples only.
Test Conditions
Patient Lead or Combination Measured
System Power
Grounding/PE
RA-LA
LA-LL
LL-RA
Mains on
open
RF on
open