GE H
EALTHCARE
DRAFT
V
OLUSON
E8 / V
OLUSON
E6
D
IRECTION
KTD102576, R
EVISION
7
DRAFT (A
UGUST
23, 2012)
S
ERVICE
M
ANUAL
Chapter 7 - Diagnostics/Troubleshooting
7-15
7-5-4
CSD: Configuration
7-5-4-1
How to configure the Service Platform
1.) Ask the hospital’s network Administrator for the Proxy Server Address and the Proxy Server Port.
2.) Select the
Configuration
page, then double-click “InSite ExC Agent Configuration”.
3.) Fill out all
bold
stated “mandatory” fields.
4.) Enter Proxy Server Address and Proxy Server Port.
5.) Click the SUBMIT CHANGES button and then close the service page.
NOTICE
!! NOTICE:
If a Proxy server is available, the
information MUST be properly entered, otherwise remote control
does not work
. There is no possibility to detect proxy server information automatically.
Figure 7-17 Common Serive Desktop - Configuration
enter Proxy Server
Information
Submit Changes
Example shows
fictional data!
NOTE:
“Device” and “Serial Number”
is
KE810026
(prefix KE8 + system´s serial
number
without
prefix-letter).
NOTE:
Bold stated fields
are mandatory!
(must be filled out).