4 Operation
4.5 Procedures after scanning
54
Typhoon
Operating Instructions
28-9623-56
AA
Turn off Typhoon Scanner
Control software and
instrument
1
Select
File:Exit
to exit the Typhoon Scanner Control software.
If you want to scan again, start the Typhoon Scanner Control software as described
(
Section 4.2 Starting the instrument and Typhoon Scanner Control software‚ on
2
Press the
Power
switch on the lower right side of the
Typhoon
instrument to the off
position to turn off the
Typhoon
instrument. The power indicator light turns off.
Note:
You can continue to use the computer after you turn off the Typhoon
instrument. However, if you want to use the instrument again, you must
follow the procedures described in Section 4.2 Starting the instrument and
Typhoon Scanner Control software‚ on page 28.
3
Save and close any open files in the computer.
4
Close all the running applications in the computer.
5
Turn off the computer.
NOTICE
Never turn off the
Typhoon
instrument while scanning. The internal
mechanism of the instrument can be severely damaged.