Typhoon
Operating Instructions
28-9623-56
AA
35
Operation 4
Preparations before start 4.3
c)
Provide enough time to register the contamination as an image on the screen.
d) After storage, scan the screen (see
Section 4.4 Performing a scanning‚ on
) and examine the image.
•
If contamination is still present, clean and erase the screen again and re-
check for contamination.
•
If no contamination appear on the image the screens is ready to use.
Storage phosphor screen
exposure
Note:
This section is valid only when performing
Storage phosphor screen
autoradioagraphy.
Recommendations for storage phosphor screen exposures
•
Use an Exposure cassette to expose the screen to the sample.
•
Expose thick samples, such as wet gels or TLC plates which are too thick to fit in
an Exposure cassette, together with a screen in a light-tight drawer or similar.
•
To expose multiple samples on the same screen, make sure the samples have the
same thickness.
•
It is important to place the screen on the sample correctly the first time. Adjusting
the position of the screen can result in a double exposure.
•
If you must adjust the position, remove the screen and erase it.
Exposure procedure
1
Select the Exposure cassette size and style that matches the Storage phosphor
screen you want to use with the sample, see
.
2
Use a damp cloth to clean the grid surface inside the Exposure cassette and
remove any radioactive contamination.
NOTICE
Do not put uncovered wet gels in an Exposure cassette.
Содержание Typhoon 9200
Страница 2: ...GE Healthcare Typhoon Operating Instructions Original instructions...
Страница 3: ......
Страница 13: ...1 Introduction 1 4 Control software 12 Typhoon Operating Instructions 28 9623 56 AA...
Страница 27: ...3 Installation 3 5 Spare parts and accessories 26 Typhoon Operating Instructions 28 9623 56 AA...
Страница 75: ...Appendix A Accessories A 4 Beamsplitters 74 Typhoon Operating Instructions 28 9623 56 AA...
Страница 76: ......