4 Operation
4.5 Procedures after scanning
52
Typhoon
Operating Instructions
28-9623-56
AA
5
A
More Info
button appears in the
ImageQuant Preview
window when the
Typhoon instrument reports errors. Click the
More Info
button to view the error
message.
•
If the error is a nonfatal error, the instrument continues to scan, and Scanner
Control saves the file when the scan is finished. Check the image in
ImageQuant to make sure the data are not corrupted.
•
If the error is a fatal error, the instrument aborts the scan, and Scanner Control
saves the data collected before the fatal error occurred. Check the image in
analysis software. Depending on where the fatal error occurred in the scan,
you might be able to use the data.
6
Check the preview image for saturated data.
While the instrument is scanning, an image appears showing the part of the
sample that has been scanned in the
ImageQuant Preview
window. Saturated
data appear in red in the image. It is not possible to perform quantitative analysis
on the portions of the image that are saturated. A saturated image can be used
only for viewing the shapes and positions of the spots on the sample.
4.5 Procedures after scanning
After scanning procedures
After scanning
1
Typhoon Scanner Control
software
saves the image using the file name you
selected in the
Save As
window.
2
The
Scan in Progres
s window displays a
Complete
message.
Saturated data appear in red in the Scan in Progress window. If the image appears
too saturated, you might not be able to analyze the image correctly.
Note:
Before repeating the scan, verify in analyzing software that the image is not
too saturated to analyze.
Note:
If scanning a
Storage phosphor screen
use a clean
Storage phosphor screen
to re-expose the original sample or to a new sample. since the scanning
process destroys the signal on the used
Storage phosphor screen
, To avoid
saturation of the data, expose the screen for a shorter time.
3
The Scan indicator light on the
Typhoon
instrument turns off.
4
Open the sample lid on the instrument.
5
Remove the fluorescent sample or Storage phosphor screen.
Storage phosphor screens:
Содержание Typhoon 9200
Страница 2: ...GE Healthcare Typhoon Operating Instructions Original instructions...
Страница 3: ......
Страница 13: ...1 Introduction 1 4 Control software 12 Typhoon Operating Instructions 28 9623 56 AA...
Страница 27: ...3 Installation 3 5 Spare parts and accessories 26 Typhoon Operating Instructions 28 9623 56 AA...
Страница 75: ...Appendix A Accessories A 4 Beamsplitters 74 Typhoon Operating Instructions 28 9623 56 AA...
Страница 76: ......