Maintenance 5
Replacement of fuses 5.7
Changing fuses in the Typhoon
instrument
You can use the Typhoon instrument with either a 220–240 V (230 V) power source or a
100–120 V (115 V) power source. The power supply in the instrument switches to the
correct voltage automatically. The fuse box is located to the right of the power switch
(
Typhoon instrument‚ on page 10
).
To change a fuse in the Typhoon instrument, follow the instruction below.
1
Turn off the Typhoon instrument and disconnect the power cord.
2
Place the end of a small flathead screwdriver into the groove (
A
) to the left
of the fuse box and pry to open the fuse box.
Figure 5-1.
Opening the fuse box of the Typhoon instrument.
3
Place the screwdriver in the groove (
B
) to the left of the fuse holder
C
) and pry out the fuse holder. When the holder is loose, pull it out of the
fuse box.
4
One or both fuses might be blown. Replace the blown fuse(s) with new fuse(s) of the
same type and rating. Fuse specifications are listed in
and on the label
located on the underside of the sample lid.
5
Insert the fuse holder into the fuse box.
6
Snap the cover of the fuse box back into place.
7
Reconnect the power cord and turn on the Typhoon instrument.
A
C
B
Содержание Typhoon 9200
Страница 2: ...GE Healthcare Typhoon Operating Instructions Original instructions...
Страница 3: ......
Страница 13: ...1 Introduction 1 4 Control software 12 Typhoon Operating Instructions 28 9623 56 AA...
Страница 27: ...3 Installation 3 5 Spare parts and accessories 26 Typhoon Operating Instructions 28 9623 56 AA...
Страница 75: ...Appendix A Accessories A 4 Beamsplitters 74 Typhoon Operating Instructions 28 9623 56 AA...
Страница 76: ......