4 Operation
4.3 Preparations before start
38
Typhoon
Operating Instructions
28-9623-56
AA
6
Place an
unmounted
screen (A) in an Exposure cassette (B).
a) The white side of the screen should face down, toward the sample (C).
b) Gently lower the screen into place over the grid area in the cassette. The
screen should be centered over the grid area and should lie straight and flat in
the cassette
c)
Close the Exposure cassette and press the lid shut until the lock clicks (D).
Sample Storage phosphor
screen exposure
Note:
This section is valid only when performing Storage phosphor screen
autoradiography.
1
Perform Storage phosphor screen autoradiography at room temperature.
2
Estimate approximately one-tenth of the exposure time of an x-ray film when
exposing Storage phosphor screens.
J
I
H
G
F
E
D
C
B
1
2
A
3
4
5
6
7
8
9
10
11
12
13
Exposure Cassette
C
B
A
D
NOTICE
Condensation can destroy the screen. If you are exposing a screen to a
frozen sample, place the screen in a sealed, dry environment, such as a
sealed bag. After exposure, allow the screen to come to room
temperature before removing it from the bag and scanning.
Содержание Typhoon 9200
Страница 2: ...GE Healthcare Typhoon Operating Instructions Original instructions...
Страница 3: ......
Страница 13: ...1 Introduction 1 4 Control software 12 Typhoon Operating Instructions 28 9623 56 AA...
Страница 27: ...3 Installation 3 5 Spare parts and accessories 26 Typhoon Operating Instructions 28 9623 56 AA...
Страница 75: ...Appendix A Accessories A 4 Beamsplitters 74 Typhoon Operating Instructions 28 9623 56 AA...
Страница 76: ......