4 Operation
4.2 Starting the instrument and Typhoon Scanner Control software
28
Typhoon
Operating Instructions
28-9623-56
AA
Principle
Storage phosphor screen scanning
Performing Storage phosphor screen autoradiography is a two step procedure, first the
Storage phosphor screen is exposed to the sample inside an Exposure cassette and,
secondly, the Storage phosphor screen is scanned in the instrument. For information of
available Storage phosphor screens and Exposure cassettes, see
Fluorescent and chemiluminescent scanning
When scanning a fluorescent or a chemiluminescent sample, the sample is placed
inside the instrument and is scanned directly.
For fluorescent scanning there are a standard set of emission filters available, see
. For fluorescent and chemiluminescent scanning beam splitters are
available, see
.
4.2 Starting the instrument and Typhoon Scanner Control
software
1
Press the
Power
switch (A) on the lower right side of the instrument.
Typhoon instruments 9200 and 9210.
Typhoon instruments 9400 and 9410 with Blue laser module.
2
The
Power
light on the front of the instrument (B) turns on and remains red during
the self-test sequence, which takes a few seconds.
A
B
A
B
C
Содержание Typhoon 9200
Страница 2: ...GE Healthcare Typhoon Operating Instructions Original instructions...
Страница 3: ......
Страница 13: ...1 Introduction 1 4 Control software 12 Typhoon Operating Instructions 28 9623 56 AA...
Страница 27: ...3 Installation 3 5 Spare parts and accessories 26 Typhoon Operating Instructions 28 9623 56 AA...
Страница 75: ...Appendix A Accessories A 4 Beamsplitters 74 Typhoon Operating Instructions 28 9623 56 AA...
Страница 76: ......