2026265-075C
Solar 8000M/i patient monitor
B-5
Appendix B – Electromagnetic compatibility:
Electromagnetic compatibility (EMC)
Recommended separation distances
The table below provides the recommended separation distances (in
meters) between portable and mobile RF communications equipment and
the Solar 8000i.
The Solar 8000M/i is intended for use in the electromagnetic
environment on which radiated RF disturbances are controlled. The
customer or the user of the Solar 8000M/i can help prevent
electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment
(transmitters) and the Solar 8000M/i as recommended below, according
to the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the
recommended separation distance [
d
] in meters (m) can be estimated
using the equitation applicable to the frequency of the transmitter,
where
P
is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer.
NOTE
These guidelines may not apply in all instances. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Separation distance in meters (m) according to frequency of transmitter
Rated maximum output
power of transmitter in
Watts
150 kHz to 80 MHz
a
d
= 1.2
80 MHz to 800 MHz
a
d
= 1.2
800 MHz to 2.5 GHz
a
d
= 2.3
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
a
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
P
P
P
Содержание Solar 8000M/i
Страница 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Страница 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Страница 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Страница 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Страница 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Страница 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Страница 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Страница 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Страница 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Страница 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Страница 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Страница 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Страница 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Страница 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Страница 180: ...A 28 Solar 8000M i patient monitor 2026265 075C Appendix A Technical specifications Solar 8000M i display ...
Страница 181: ...2026265 075C Solar 8000M i patient monitor 8000i Patient Monitor B 1 B Appendix B Electromagnetic compatibility ...
Страница 191: ......