2026265-075C
Solar 8000M/i patient monitor
4-19
Maintenance and checkout:
Checkout procedure
5. The following step does not apply for a Tram-rac 2 housing. Check
the analog output connector (yellow) using an oscilloscope. Observe a
signal at the appropriate pins found in the next table. The output
signal is dependent upon which Tram and input module functions
are activated at the monitor.
12
6
1
11
10
7
9
8
2
5
4
3
REAR VIEW
TRAM-RAC 4A
ANALOG OUTPUT
CONNECTOR
FRONT VIEW
TRAM-RAC 3
ANALOG OUTPUT
CONNECTOR
8
7
6
5
4
3
2
1
14
13
12
11
10
9
15
019A
Analog output signals
Pins for
D-type
connector
Pins for round
connector
Signal source
Tram-rac 4A
bezel number for
BP output
Pin 1
Pin 8
Signal GND for Tram Waveforms
–
Pin 2
Pin 2
Trace I (ECG II
1
)
Tram
1
Pin3
Pin 6
Tram BP3 or SPO
2
Value
Tram
1
Pin 4
–
Reserved for Future Use
–
Pin 5
Pin 4
Tram ART 1 or BP1
Tram
Pin 6
Pin 9
Slot 3 Series 7000 Waveform A (Right Side or Module)
Parameter 6
Pin 7
Pin 11
Slot 4 Series 7000 Waveform A (Right Side or Module)
Parameter 8
Pin 8
Pin 8
Signal GND for Series 7000 Waveforms
–
Pin 9
Pin 1
Tram ECG II
Tram
1
Pin 10
Pin 3
Tram ECG V
Tram
1
Pin 11
Pin 7
Tram BP4 or RESP
Tram
1
Pin 12
–
Reserved for Future Use
–
Pin 13
Pin 5
Tram BP2 or SPO
2
Waveform
Tram
Pin 14
Pin 10
Slot 3 Series 7000 Waveform B (Left Side or Module)
Parameter 5
Pin 15
–
Slot 4 Series 7000 Waveform B (Left Side or Module)
Parameter 7
1. The top displayed trace on the monitor is present unless AVR, AVL, or AVF leads are used,
then lead II is output.
Содержание Solar 8000M/i
Страница 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Страница 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Страница 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Страница 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Страница 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Страница 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Страница 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Страница 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Страница 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Страница 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Страница 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Страница 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Страница 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Страница 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Страница 180: ...A 28 Solar 8000M i patient monitor 2026265 075C Appendix A Technical specifications Solar 8000M i display ...
Страница 181: ...2026265 075C Solar 8000M i patient monitor 8000i Patient Monitor B 1 B Appendix B Electromagnetic compatibility ...
Страница 191: ......