Solar 8000M/i patient monitor
5-19
Troubleshooting:
Troubleshooting software updates
Troubleshooting software updates
The following is a list of problems commonly encountered during a
software update with their solutions.
Problem
Possible reason and solution
Monitor appears ‘locked up’
during a network download.
CAUTION
Do not power cycle or reboot the monitor if downloading the
Boot Code is proceeding normally. The monitor will be
rendered useless.
If the packet or byte numbers stop advancing for at least two minutes, do the following:
1. Check that all cables are properly connected.
2. At the patient monitor:
Hold down
NBP Go/Stop
and
Zero All
.
Press and release the
TRIM KNOB
control.
Keep holding
NBP Go/Stop
and
Zero All
until the Boot Code information appears on the display.
3. Repeat the software update procedure for the aborted file from the beginning.
Software revision window
does not list part numbers.
If the part numbers are not listed for the monitor interfaces in the software revision window, the software
update has not been activated. Power cycle the monitor and view the software revisions window again.
If the part numbers are still missing, repeat the update procedure for each missing file.
Waveforms do not appear
at the central station.
If communication is corrupted, do the following:
1. Check all cables for a good connection.
2. Ensure that the central station software is correct.
3. Ensure Ethernet addresses have been programed correctly. Refer to the appropriate service manual.
4. Ensure the Ethernet address has been programmed correctly at the patient monitor:
Hold down
NBP Go/Stop
and
Zero All
.
Press and release the
TRIM KNOB
control.
Keep holding
NBP Go/Stop
and
Zero All
until the Boot Code information appears on the display.
The Ethernet address displays in the Boot Code banner information.
5. If the Ethernet address needs to be changed in Boot Code, a unique password is required to access
Change Ethernet Address
in the
Options Menu
. Fax a password request to GE
Customer
Relationship Center at 877-633-8181 to obtain a password. You will need to provide your product
serial number and Ethernet address. (The Ethernet address displays in the Boot Code banner
information.)
Содержание Solar 8000M/i
Страница 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Страница 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Страница 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Страница 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Страница 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Страница 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Страница 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Страница 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Страница 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Страница 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Страница 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Страница 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Страница 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Страница 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Страница 180: ...A 28 Solar 8000M i patient monitor 2026265 075C Appendix A Technical specifications Solar 8000M i display ...
Страница 181: ...2026265 075C Solar 8000M i patient monitor 8000i Patient Monitor B 1 B Appendix B Electromagnetic compatibility ...
Страница 191: ......