2026265-075C
Solar 8000M/i patient monitor
7-7
Field replaceable units:
Disassembly
Disassembly
Guidelines
Follow these guidelines when disassembling the monitor. You will need a
standard set of hand tools including a #1 Phillips head screwdriver.
WARNING
Repair to the FRU level. Field repairs are recommended
to the field replaceable unit (FRU) only. Attempting a
field repair on a pcb or a factory sealed component or
assembly could jeopardize the safe and effective
operation of the patient monitor.
Electrostatic discharge (ESD) precautions
All external connectors of the Solar 8000M/i patient monitor are
designed with protection from ESD damage. However if the module
requires service, exposed components and assemblies inside are
susceptible to ESD damage. This includes human hands, non-ESD
protected work stations or improperly grounded test equipment.
The following guidelines may not guaranty a 100% static-free
workstation, but can greatly reduce the potential for failure of any
electronic assemblies being serviced:
Discharge any static charge you may have built up before handling
semiconductors or assemblies containing semiconductors.
A grounded, antistatic wristband (3M part number 2046 or
equivalent) or heel strap should be worn at all times while handling
or repairing assemblies containing semiconductors.
Use properly grounded soldering and test equipment.
Use a static-free work surface (3M part number 8210 or equivalent)
while handling or working on assemblies containing semiconductors.
Do not
remove semiconductors or assemblies containing
semiconductors from antistatic containers (Velo-stat bags) until
absolutely necessary.
Make sure power to an assembly is turned off before removing or
inserting a semiconductor.
Do not
slide semiconductors or electrical/electronic assemblies across
any surface.
Do not
touch semiconductor leads unless absolutely necessary.
Semiconductors and electronic assemblies should be stored only in
antistatic bags or boxes.
Handle all PCB assemblies by their edges.
Do not
flex or twist a circuit board.
Содержание Solar 8000M/i
Страница 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Страница 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Страница 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Страница 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Страница 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Страница 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Страница 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Страница 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Страница 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Страница 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Страница 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Страница 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Страница 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Страница 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Страница 180: ...A 28 Solar 8000M i patient monitor 2026265 075C Appendix A Technical specifications Solar 8000M i display ...
Страница 181: ...2026265 075C Solar 8000M i patient monitor 8000i Patient Monitor B 1 B Appendix B Electromagnetic compatibility ...
Страница 191: ......