4-30
Solar 8000M/i patient monitor
2026265-075C
Maintenance and checkout:
Checkout procedure
The pump shuts off at about 250 mmHg, and the pressure drops
slowly to about 240 mmHg before stabilizing. The second line of text
on the
CAL GAIN
menu item changes from
INFLATING
to
HOLDING
NOTE
If the pressure continues to drop at a rate of 1 mmHg or more for
every five seconds, there is a leak in the NBP plumbing. If there is a
leak in the NBP plumbing, correct the problem and restart this
calibration procedure.
10. Select
ENTER CAL PRESSURE
.
11. An
ENTER CAL PRESSURE
popup window appears. Use the
Trim
Knob
control to select a pressure value that is 1 mmHg lower than
the current manometer reading.
12. When the manometer falls to exactly the value that you selected in
the popup window, press the
Trim Knob
control to enter the value.
Select
PREVIOUS MENU
to return to the
MAIN MENU
button.
13. Select
CHECK CAL OFF
.
14. Select
START
.
15. The text on the menu item changes from
CHECK CAL OFF
to
CHECK CAL IN PROGRESS
. Make sure that the pressure
readings (shown as
CUFF
in the NBP parameter box) on the monitor
display and manometer are equal (± 1 mmHg) for at least one full
minute.
16. Select
CHECK CAL IN PROGRESS
.
17. Select
STOP
. The module then releases pressure in the bulb or cuff.
18. Remove the cuff and manometer from the acquisition device.
Defib/sync and analog output test
Patient Data Module
1. Connect the unterminated cable assembly to the
Defib/Sync
connector on the front of the Patient Data Module.
2. Set up the simulator:
Heart rate = 80 bpm
Amplitude to 1.0 mV
3. Test the ECG, Arterial BP, and Marker Out signals from the
DEFIB
SYNC
connector using an oscilloscope to connect to the appropriate
wire on the test cable (shown below). They should resemble the
waveforms in the following figures. Note that there are two Marker
Out traces shown. The top trace shows the frequency of the pulses;
the bottom trace shows the pulse width.
Содержание Solar 8000M/i
Страница 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Страница 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Страница 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Страница 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Страница 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Страница 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Страница 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Страница 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Страница 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Страница 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Страница 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Страница 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Страница 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Страница 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Страница 180: ...A 28 Solar 8000M i patient monitor 2026265 075C Appendix A Technical specifications Solar 8000M i display ...
Страница 181: ...2026265 075C Solar 8000M i patient monitor 8000i Patient Monitor B 1 B Appendix B Electromagnetic compatibility ...
Страница 191: ......