4-24
Solar 8000M/i patient monitor
2026265-075C
Maintenance and checkout:
Checkout procedure
Invasive blood pressure functions
1. Set up simulator:
Go to
BP
functions, select
BP1
, then
0
.
Set output to
0 mmHg
.
2. Connect the IP simulator cable from the
BP1
connector of the
simulator to the corresponding
IP
connection of the acquisition
device.
3. Observe an
ART1
label and graticules on the patient monitor.
4. Zero the
ART1
waveform by pressing the Zero All button on the
Patient Data Module, Solar keypad or remote.
5. Set the simulator to output
240 mmHg
. (
BP1
>
STAT
>
UP
>
RUN
)
6. Observe a reading of
240
± 4 mmHg on the patient monitor.
7. Set the simulator to BP dynamic waveform output. (
BP1
>
DYNA
>
RUN
)
8. Set the patient monitor
ART1 SCALES
to
AUTO
.
9. Observe a distortion-free waveform and a blood pressure reading of
approximately 120/80 (93) on the patient monitor.
10. Repeat these tests with each
IP/BP
connector selecting
BP2
,
BP3
,
etc.
NOTE
Use the Dual IP Adapter Cable to test the third and fourth IP
functionality for each IP/BP connection.
Temperature functions
1. Go to
TEMP
functions on the simulator and set a 37°C temperature
output.
2. Connect a temperature sensor adaptor to the
TEMP/CO
connector of
the acquisition device.
3. Set the switch on the adaptor to the
400
position.
4. Connect the temperature simulator cable from the temperature
output connector of simulator to the
T1
connector of the temperature
sensor adaptor.
5. Observe that the
T1
reading on the patient monitor is between 36.6
and 37.4.
6. Move the cable from the
T1
connector of the temperature sensor
adaptor to the
T2
connector.
7. Observe that the
T2
reading appears on the patient monitor is
between 36.6 and 37.4.
8. Disconnect the temperature sensor adaptor and temperature cable
from the acquisition device and the simulator.
Содержание Solar 8000M/i
Страница 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Страница 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Страница 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Страница 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Страница 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Страница 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Страница 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Страница 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Страница 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Страница 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Страница 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Страница 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Страница 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Страница 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Страница 180: ...A 28 Solar 8000M i patient monitor 2026265 075C Appendix A Technical specifications Solar 8000M i display ...
Страница 181: ...2026265 075C Solar 8000M i patient monitor 8000i Patient Monitor B 1 B Appendix B Electromagnetic compatibility ...
Страница 191: ......