2026265-075C
Solar 8000M/i patient monitor
4-17
Maintenance and checkout:
Checkout procedure
Set up
Complete the following steps in the order presented. Failure to attain
any of the listed results indicates a malfunction.
1. Confirm that all components of the monitoring system are correctly
connected as described in Chapter 3,
For Patient Data Module, connect using the ePort host interface
cable.
For Tram, place the Tram module into the top two slots of the
Tram-rac housing.
2. Verify that the power indicator illuminates on the acquisition device.
3. Using the patient simulator and patient cables, configure the
monitor display with as many waveforms as possible. Refer to the
appropriate patient monitor operator's manual, if necessary.
4. The waveforms should look clean (no noise).
Temperature test:
700/400 series dual temperature adapter (402015-004)
Temperature simulator cable for use with Marq III
Cardiac output test:
Cardiac output simulator II (900028-001) (no longer available for ordering)
Cardiac output cable adapter (700092-001)
Sp02 test:
Masimo SpO2 Test Kit includes Masimo Tester and SpO2 Sensor Adapter Cable
(2021087-001)
Nellcor pulse oximeter tester model SRC-2 (2007650-001)
Dual BP adapter cable (2005772-001)
Nellcor OxiMax pulse oximeter functional tester model SRC-MAX (2007650-002)
NBP test:
NBP cuff coupling (400787-001)
NBP hose coupling (46100-0020)
NBP tee (4745-101)
NBP tubing 2 feet (401582-001)
Manometer, digital or mercury (Meriam Instrument Smart Manometer model 350
DM2000 or equivalent)
NBP tube (414873-001)
NBP cuff (9461-301)
Pipe, PVC
DEFIB sync test:
Unterminated cable assembly (2017842-001)
Oscilloscope
Содержание Solar 8000M/i
Страница 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Страница 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Страница 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Страница 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Страница 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Страница 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Страница 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Страница 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Страница 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Страница 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Страница 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Страница 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Страница 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Страница 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Страница 180: ...A 28 Solar 8000M i patient monitor 2026265 075C Appendix A Technical specifications Solar 8000M i display ...
Страница 181: ...2026265 075C Solar 8000M i patient monitor 8000i Patient Monitor B 1 B Appendix B Electromagnetic compatibility ...
Страница 191: ......